Chapter 8 Troubleshooting
2 Image Output to a PC Monitor
- 34 -
Trouble Check point
The contrast is low.
Check that the tone is adjusted correctly with the application software.
Refer to the instruction manual for the application software.
Check that the black level is set appropriately with the application
software.
Refer to the instruction manual for the application software.
Check that the size of the field diaphragm is adjusted so that it
circumscribes the view field.
Check that the size of the aperture diaphragm is adjusted to
approximately 70% to 80% of the numerical aperture of the objective
lens.
Check that an appropriate filter is used.
When using a phase contrast microscope or an interference microscope
(two luminous fluxes or multiple luminous fluxes), insert the specified
filter. With a black and white image, contrast is improved by inserting a
green interference filter (GIF). Contrast in color images is enhanced by
inserting a filter having a complementary color to that of the specimen.
Check that the ambient light around the microscope does not affect.
If the ambient light affects, darken the room or attach a cap to the
binocular eyepiece.
Check that the contrast of the specimen is sufficient.