EasyManua.ls Logo

NIPRO Diagnostics True Metrix Pro - Test Strip Design and Technology

NIPRO Diagnostics True Metrix Pro
63 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TRUE METRIX
®
PRO System Comprehensive Resource Guide (GDH-FAD Enzyme) Test Strip Design 10
Test Strip Design
TRUE METRIX
®
PRO Test Strips –
Featuring TRIPLE SENSE TECHNOLOGY
DETECT
Sample Environment
An electrode pair on the test
strip detects hematocrit
An internal unit in the meter
detects temperature
Sample Size
System triggers proper fill
detection eliminating errors
due to improper sample size
Control
Detects control solution and automatically identifies
results in memory for accurate data trending
ANALYZE
Environmental and
Physiological Factors
Every system analyzes the testing
environment for hematocrit and
temperature, which could impact
or influence the accuracy of results
Testing outside of specified operating temperature
range is marked by an error code – eliminating
incorrect results
CORRECT
Hematocrit/Temperature
Variables
Employs a Proprietary
Dynamic Adaptive
Algorithm, which incorporates:
– Active hematocrit correction
– Temperature compensation
Combined, this algorithm corrects results based
on the testing environment
• Ensures proven accuracy and confidence in results
No Coding Technology eliminates the need for coding
of the meter.
• Code is assigned during manufacturing process
• Code is printed onto contacts on end of test strip
• Meter reads code on insertion
Advanced technologies – the meter, a complex
algorithm, chemistry and electrodes – on the test
strip work together as part of the TRUE METRIX®
PRO system to produce accurate results.
Featuring TRIPLE SENSE TECHNOLOGY™, the
system provides proven accuracy and confidence in
results.

Related product manuals