TRUE METRIX
®
Control Solution
• Write the date first opened on control
solution bottle label. Discard if either 3
months after first opening or after date
printed next to EXP on vial label has passed,
whichever comes first.
• After use, wipe bottle tip using a clean, dry
cloth and recap tightly.
• Store at temperatures between, 36°F-86°F.
DO NOT FREEZE.
System Storage
TRUE METRIX
®
PRO System
• Store the system (meter, control solution,
test strips) in carrying case to protect from
liquids, dust and dirt.
• Do not keep meter in an area where it may
be crushed (i.e. back pocket, drawer,
bottom of bag, etc.).
• Store the system in a dry place
at temperatures between, 40°F-86°F.
DO NOT FREEZE.
• Allow System to sit at room
temperature for 10 minutes
before testing
TRUE METRIX
®
PRO Test Strips
• Store test strips in original vial only. Do not
transfer old test strips into new vial or
store test strips outside of vial.
• Write the date first opened on test strip
vial. Discard unused test strips from vial
if either 4 months after first opening or
after date printed next to EXP on vial label
has passed, whichever comes first. Use of
test strips past expiration dates may give
incorrect results. Discard out of date test
strips and use new test strips.
• Close test strip vial immediately
after removing one test strip.
• Store in a dry place at temperatures
between, 40°F – 86°F. DO NOT FREEZE.
• Do not reuse test strips.
• Do not bend, cut or alter test strips in any
way.
TRUE METRIX
®
PRO System Comprehensive Resource Guide (GDH-FAD Enzyme) System Storage 33