TRUE METRIX
®
PRO System Comprehensive Resource Guide (GDH-FAD Enzyme) System Specifications 4
System Specifications
Meter Specifications
Result Range: 20-600 mg/dL
Sample Size: Minimum 0.5 microliter (0.5 μL)
Sample Type: Fresh capillary whole blood from the
finger or forearm, venous whole blood collected
in sodium heparin blood collection tubes only, or
control solution
Test Time: Results in as little as 4 seconds
Result Value: Plasma equivalent values
Assay Method: Amperometric
Reference Method: Yellow Springs Instrument
Glucose oxidase reagent
Power Supply: One 3V lithium battery #CR2032
(non-rechargeable).
Battery Life: Approximately 1,000 tests or 1 year
Automatic Shut-Off: After two minutes of non-use
Weight: 1.66 ounces
Size: 3.44" x 2.16" x 0.69"
Memory Size: 500 glucose and control results
Operating Conditions
System Operating Range (meter & test strips):
Relative Humidity: 10-90% (Non-condensing)
Temperature: 41°F-104°F
Hematocrit: 20-70%
Altitude: Up to and including 10,200 feet
Note: Use within specified environmental
conditions only.
Chemical Composition
TRUE METRIX
®
PRO Test Strips: Glucose
dehydrogenase-Flavin adenine dinucleotide
(GDH-FAD) (Aspergillus sp.), mediators, buffers
and stabilizers.
TRUE METRIX
®
Control Solution: Water, d-glucose,
buffers, viscosity enhancing agent, salts, dye
and preservatives.
Note: Material Safety Data Sheets can be found in
the Appendix.