38
Technical Information
Equipment Response Time
Example - SpO
2
Exponential Averaging
SpO
2
decreases 0.75% per second (7.5% over 10 seconds)
Pulse Rate = 75 BPM
Specific to this example:
The response of the 4-beat average is 1.5 seconds.
The response of the 8-beat average is 3 seconds.
Pulse Rate Values Response Latency
Standard/Fast Averaged Pulse Rate 4 beat exponential 2 beats
Extended Averaged Pulse Rate 8 beat exponential 2 beats