38
Testing: Blood and Quality Control Solutions
NOTE: Cleaning of the puncture site is important. Sweat left
on the ngers may cause falsely high lactate test results.
5. Holding hand downward, massage nger with thumb
toward tip to stimulate blood ow.
6. Use a single-use, disposable safety lancet to puncture
the nger. To reduce the risk of pre-analytical error,
facilities should consider using a 21-gauge lancet when
collecting capillary samples.
7. Wipe off the rst drop with a clean tissue. The rst drop
may be contaminated.
8. Squeeze the nger to form a second drop of blood. To
avoid collection of interstitial uid as well as blood, do
not squeeze vigorously.