39
Testing: Blood and Quality Control Solutions
9. When the second blood drop appears, touch the end of
the test strip to the blood drop until the well of test strip
is full and the meter beeps.
NOTE: The on-screen Blood Drop ashes on and off repeat-
edly until sufcient blood has been added to the test strip.
10. If subject has to immediately return to exercise and
puncture site is still bleeding, apply light pressure with
a cotton pad or tissue until the bleeding stops or cover
with a bandage.
11. The Lactate result is available on-screen in 13 seconds.
(See pages 42 and 43.)
12. Remove the strip manually or use the ejector button
on the back of the meter to eject the strip directly into a
biohazard container.