NOTE: To properly clean and disinfect the Meter, steps 1
to 5 should be performed together before testing on each
patient.
1. Clean the Meter.
• Wipe the external surface of the Meter thoroughly with a fresh
Germicidal Wipe.
• Discard the used wipe per Step 4.
2. Disinfect the Meter.
• Remove another fresh Clorox Germicidal Wipe from the
canister. Thoroughly wipe the top, bottom, left, and right
sides of the Meter avoiding the Test Strip port by wiping
the surface a minimum of 3 times horizontally followed
by 3 times vertically.