E3 Used Strip Error
The test strip was previously used.
Action: Repeat the test with a new test strip.
If the error persists, perform the test using an
alternate test strip vial or alternate method.
E4 Short Sample Error
An insufcient sample volume (Control or blood)
was drawn into the test strip.
Action: Repeat the test with a new test strip.
If the error persists, perform the test using an
alternate test strip vial or alternate method.