User Manual TXLD0012
P/N T2-UM-E-ENG Rev 10 Novoxel.com Page | 65
Immunity Test IEC 60601 Level Compliance Level
Electromagnetic Environment -
Guidance
Voltage dips and
interruptions on
power supply input
lines
IEC 61000-4-11
0% U
T
for 0.5
cycle
0% U
T
for 1 cycle
70% U
T
for 25/30
cycles
0% U
T
for 250/
300 cycles
0% U
T
for 0.5
cycle
0% U
T
for 1 cycle
70% U
T
for 25/30
cycles
0% U
T
for 250/300
cycles
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the equipment
requires continued operation during
power mains interruptions, it is
recommended that the equipment be
powered from an uninterruptible power
supply or a battery.
Power frequency
magnetic field,
IEC 61000-4-8
30 A/m 30 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Immunity Test IEC 60601 Level Compliance Level
IEC 61000-4-6
Conducted RF
3 Vrms
150 kHz to 80 MHz
6 Vrms in ISM bands (6.765 MHz to 6.795 MHz;
13.553 MHz to 13.567 MHz; 26.957 MHz to
27.283 MHz; and 40.66 MHz to 40.70 MHz) and
amateur bands (1.8 MHz to 2.0 MHz, 3.5 MHz to
4.0 MHz, 5.3 MHz to 5.4 MHz, 7 MHz to 7.3
MHz, 10.1 MHz to 10.15 MHz, 14 MHz to 14.2
MHz, 18.07 MHz to 18.17 MHz, 21.0 MHz to
21.4 MHz, 24.89 MHz to 24.99 MHz, 28.0 MHz
to 29.7 MHz and 50.0 MHz to 54.0 MHz).
[V] = 3 Vrms
[V] = 6 Vrms
IEC 61000-4-3
Radiated RF
10 V/m
80 MHz to 2.7 GHz
[E] = 10 V/m
Proximity fields from
RF wireless
communications
equipment.
385 MHz 27 V/m
450 28 V/m
710
9 V/m745
780
810 28 V/m