Bluetooth® Wireless Technology
The Nox A1 recorder uses Bluetooth® 2.0 wireless technology to communicate with external
Bluetooth modules.
The Bluetooth wireless technology is based on a radio link that offers fast and reliable transmission of
data. Bluetooth radio uses globally available frequency range in the industrial, scientific and medical
(ISM) band, intended to ensure communication compatibility worldwide and a fast acknowledgement
and frequency-hopping scheme to make the link robust, even in noisy radio environments. Please
refer to the Specifications” section for details on Radio Frequency (RF) specifications for the Nox A1
recorder.
The Bluetooth® word mark and logos are registered trademarks owned by the Bluetooth SIG, Inc. and
any use of such marks by Nox Medical is under license. Other trademarks and trade names are those
of their respective owners.
Electromagnetic Compatibility (EMC) Information
Caution: Exposure to radio frequency radiation.
Portable and mobile Radio Frequency (RF) communications can affect the
performance of the Nox A1 recorder.
Warning: The Nox A1 recorder(s) should not be used adjacent to, or stacked with
other equipment. If adjacent or stacked use is necessary, the device(s) should be
observed to verify normal operation in the configuration in which it will be used
and prevent abnormal operation which might cause injuries to the
operator/patient.
Warning: The use of accessories, transducers, sensors, and cables other than those
listed in this manual may result in increased emissions and/or decreased immunity
of the Nox A1 recorder and cause injuries to the operator/patient.
Warning: The Nox Sleep System may be interfered with by other equipment, even
if that equipment complies with CISPR (Special International Committee on Radio
Interference) emission requirements, causing possible patient harm
Refer to the tables below in this section for specific information regarding the Nox
A1 recorder’s compliance to the standard IEC60601-1-2: Medical electrical
equipment - Part 1-2: General requirements for basic safety and essential
performance - Collateral Standard: Electromagnetic disturbances - Requirements
and tests.