~ 20 ~
Measuring Mask Pressure
Warning: The mask pressure tubes and Nox filter tube connectors are single
patient use. Using the same mask pressure tube and filter tube connector on more
than one patient poses a risk of cross-infection.
Note: The mask pressure tube can only be connected to the pressure lock on the
Nox A1 recorder by using the Nox filter tube connector.
A mask pressure tube is used for connection to positive airway pressure (PAP) masks for measuring
mask pressure. The pressure tube connects to the pressure lock on the Nox A1 recorder via a filter
tube connector from Nox Medical.
Refer to the “Compatible Sensors and Devices” section regarding the types of mask pressure tubes
that have been validated with the Nox A1 recorder.
Measuring EEG Signals
Warning: Electrodes should only be used by or in consultation with a healthcare
provider familiar with their proper placement and use. Not using and placing the
electrodes correctly may affect recording of data, and therefore interpretation and
diagnostic.
Warning: The electrodes should be applied only to intact, clean skin (e.g. not over
open wounds, lesions, infected or inflamed areas) in order to prevent infections.
Warning: Make sure the conductive parts of electrodes and associated connectors,
including the neutral electrode, do not contact other conductive parts including
earth to prevent potential serious harm to the operator/patient.
Warning: The Nox 5 Lead EEG Electrode Cables should be properly disposed of if
they cannot be fully cleaned between uses in order to prevent the risk of cross-
infection between patients.
Warning: The Nox 5 Lead EEG Electrode Cables are not certified to be used for
electrical stimulation purposes. Using the product for electrical stimulation
purposes might create burns and cause injuries to the patient.
Warning: The Nox EEG Head Cable/Nox 5 Lead EEG Cables do not provide
protection against the effect of the discharge of a cardiac defibrillator nor against
high frequency burns. Not removing the equipment from a patient before
defibrillation may lead to the creation of high current density at the electrode
sites, causing burns and leading to possible patient harm.
Note: The Nox EEG Head Cable is available in both pediatric and adult sizes.