Cleaning / Checking the Irrigation Nozzle
English
Manual Cleaning without Cleaning Tools
_ Pre-cleaning
Remove contaminants on the surface with a cloth or soft bristle brush under running tap water.
≤ 38°C, ≥ 3.5 L/min, ≥ 15 sec.
Water quality: Drinking water
_ Cleaning
1
Prepare cleaning solution in the cleaning tank under the
following conditions (prepare the cleaning solution according to
the instructions of the detergent manufacturer).
Temperature Water quality Detergent
45°C
Distilled water
Deionized water
Demineralized
water
Neutral enzymatic
detergent for medical
purposes (pH 7.0 - 8.0)
* Detergent used for validation:
neodisher MediZym (DR. WEIGERT)
2
Immerse the product in the cleaning tank for 20 - 30 min.
3
Thoroughly rinse the dirt released in immersion as well as the
detergent under running tap water.
≤ 38°C, ≥ 3.5 L/min, ≥ 10 sec.
Water quality: Drinking water
4
Rinse the product surface with distilled, deionized or
demineralized water.
_ Drying
1
Remove moisture inside and on the surface of the product using
a dry cloth, air gun, etc.
2
Visually check that all contamination is removed.
In case there is residual contamination, repeat the procedures
of Pre-cleaning through Drying until all contamination is
removed.
Proceed to Checking (see “9 2 Checking”).