29
After the disinfection process, temperature and CO
2
controls start when the
temperature and CO
2
values are in the alarm limit.
During Auto-zero and disinfection process, alarm limit error do not appear.
If auto-zero operates when disinfection starts, auto-zero stops and auto-zero
does not operate during disinfection period. Auto-zero does not also operate
while service menu is open.
If an error occurs, please contact with an authorized Nüve agent to seek
technical help.
9.2. FUSE REPLACEMENT
The fuses shall always be replaced by the authorized personnel.
10. OPTIONS
10.1. AlerText™ GSM ALARM MODULE
GSM Module transfers data via GPRS. SMS is sent to 5 different
mobile phone numbers in case of failures. In addition to this, message
could be sent as an e- mail. The data is saved on the internal memory
of GSM module in case of failures on the communication.
There could be some errors while the messages are being delivered which
means messages have not been delivered.
10.1.1. CONNECTING THE GSM MODULE TO THE DEVICE
The optional GSM Module connection is made via NC contact output which is located
on the back side of the device. Before starting the operation, please connect the
GSM Module to the device. GSM Module has 1 adaptor with 12V/2A output, 1
removable antenna and 1 main terminal. Connection between the main terminal and
power socket should be made correctly.
Please be careful and do not connect the main terminal’s cables to the RTD
input, otherwise module will be defective and cannot be used again!