EasyManua.ls Logo

NUVE ST 30 - Operator Menu Parameters

NUVE ST 30
21 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13
Wait by pressing the encoder button.
Lift your hand when you see "oP" on the temperature
display and again press the encoder button. If the
device has a menu protection password, password
screen “oPS” will come on the temperature display.
You enter the set password to turn the encoder button
right and left. ( The password will not be asked if the
device is newly installed.) Confirm the password by
pressing the encoder button. Observe that the
parameter numbers on the temperature display change
with each pressing the encoder button. For operator
menu parameter descriptions (see section 6.1.1). You
can set the parameter values by turning the encoder
button right or left on time display.
Again press the encoder button and confirm the set
value. Press the back button to return the work screen.
6.2.2 oP OPERATOR MENU PARAMETERS
1: Recording Period: This time is recording period of temperature and error information.
2: Lid Alarm Time: Not used for ST models.
3: Lid Alarm Range: Not used for ST models.
4: Timer Set Band: When the read temperature reaches the "Set Temperature - TIMER SET
BAND" value, the time starts counting backwards.
5: Buzzer ON/OFF: The alarm sound on/off 0: OFF 1: ON
6: Date Setting - Year: Two digits are displayed the year information of date. If updating is
necessary, change.
7: Date Setting - Month: The month information of date is displayed. If updating is
necessary, change.
8: Date Setting - Day: The day information of date is displayed. If updating is necessary,
change.
9: Time Setting - Hour: The hour information of time is displayed. If updating is necessary,
change.
A: Time Setting - Minute: The minute information of time is displayed. If updating is
necessary, change.
B: Time Setting Second: The second information of time is displayed. If updating is
necessary, change.
C: Date / Time Update: 0: No change 1: Update date / time according to the entered values.
The entered values are considered as current Date / Time information when 5, 6, 7, 8, 9, A
parameters are changed and B parameter is set to 1.
D: Password: The password used to enter the operator parameters. This password used
when you want to change the set values. No password if 0 is selected.

Related product manuals