8
SECTION 5
INSTALLATION
5.1 ENVIRONMENTAL CONDITIONS
The water bath is designed to operate safely under the following conditions:
• Indoor use only
• Ambient temperature: 5°C to 40°C.
• Maximum relative humidity for temperature up to 22°C: 80%.
• Maximum altitude: 2000 m.
• Temperature for maximum performance: 15°C / 25°C.
5.2 HANDLING AND TRANSPORTATION
All handling and transportation must be carried out by using proper equipment and
experienced staff. The instrument must be supported underneath and never be turned upside
down.
5.3 UNPACKING
Open the packing cardboard box. Remove the nylon packing wrapped around the device.
The items provided with the device are listed below, please check them.
1 ea. user’s manual
1 ea. warranty certificate
1 ea. power cable
5.4 MAINS SUPPLY
The water bath requires 230 V, 50/60 Hz.
Please make sure that the supplied mains matches the required power ratings which are
written on the name plate of the instrument located at the back of the steam sterilizer.
Always plug-in the instrument to correctly grounded sockets.
A supply fitted with a circuit breaker should be used for protection against indirect
contact in case of an isolation fault.
5.5 POSITIONING
Check that no damage occurred during transportation.
Check that the positioning is suitable for the users.
Support the water bath underneath and carry it to its place carefully.
Make sure that the bench that the water bath is positioned is resistant to the weight of the
instrument and vibration free.