EasyManua.ls Logo

NUVE ST 30 - Error Codes and Explanations; Fuse Replacement

NUVE ST 30
21 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
18
9.1 ERROR CODES AND EXPLANATIONS
Err0:
Motor thermal is off. Contact to the authorized service.
Er1:
The temperature sensor endings are broken. The error code flashes on the temperature
display and an audible alarm sounds.
Er2:
An electronic failure occurs in the microprocessor. The error code flashes on the temperature
display and an audible alarm sounds.
Er3:
The temperature sensor measures a temperature higher than 147°C.The error code is shown
on the temperature display and an audible alarm sounds.
Er4:
The temperature sensor endings are connected in reverse. The error code flashes on the
temperature display and an audible alarm sounds.
EoF:
This error code appears if any probable power cut causes when the sterilization phase “EoF”
flashes and the audible alarm sounds on the temperature display.
If an error occurs, please contact to an authorized NÜVE AGENT to seek technical
help.
IN CASE OF ANY ERROR, THE PROGRAM İS STOPPED AUTOMATİCALLY AND
IMMEDIATELY.
9.2 FUSE REPLACEMENT
The fuses shall be always be replaced by the authorized personnel.

Related product manuals