17
eAdvantage User Manual (15PL1033 Rev. 7, July 2022)
Once this screen is displayed the mixer is ready for operation.
The eAdvantage® will not switch ON if any of the problems listed below
are detected:
1. O
2
Valve Failure.
2. N
2
O Valve Failure.
3. O
2
Flow Sensor Failure.
4. N
2
O Flow Sensor Failure.
5. U12 (Paw) Sensor Failure.
6. U13 (O
2
input pressure) Sensor Failure.
7. U14 (N
2
O input pressure) Sensor Failure.
Instead, a new screen will display the cause of the malfunction and
request the mixer to be checked or returned for service. At the same
time a continuous Audible Alarm will sound.
The above failures will be saved in the events log under specifi c failure
as listed in “Stored History” section
If multiple failures were detected the mixer will switch between the
messages related to the failures.
Note: Start Up Calibration Error (Startup calibration results and during
use technical failure):