eAdvantage User Manual (15PL1033 Rev. 7, July 2022)
10. TROUBLESHOOTING
WARNING
Please contact the manufacturer if a problem cannot be rectifi ed. For
the safety of the patient and the health care providers DO NOT continue
using the mixer.
Message/fault Cause Remedy
High inhalation resistance
Disconnected sensing hose
or no input gas
Check input gas valves or
sensing hose connection
External leak Input hoses are not tightened Tighten input hoses
Extremely high
inhalation resistance
Device is OFF
Remove mask from face and
ensure the device is ON
No input gas
No gas or input gas pressure
below 20 PSI
Change the gas cylinder
Low input gas Low gas below 40 PSI Change the gas cylinder
Battery discharges
quickly
No proper charging/faulty
battery
Charge battery as per
instructions/replace battery
eAdvantage® cannot
be switched ON
Battery empty/no power
supply connected/defective
Change battery/connect
power supply/ send to
O-Two™ for repair or service