eAdvantage User Manual (15PL1033 Rev. 7, July 2022)
The eAdvantage® is intended for use in the electromagnetic environment
in which radiated RF disturbances are controlled. The user can help
prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment
(including peripherals such as antenna cables and external antennas)
and the eAdvantage®.
Electromagnetic Immunity
Immunity test Immunity test level
Electrostatic discharge
(ESD) IEC61000-4-2
± 8 kV contact
± 2, ±4, ±8, ±15 kV air
Electrical fast transient/
Burst IEC61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines
Surge IEC61000-4-5
±0.5, ±1 kV line to line;
±0.5, ±1 kV, ± 2 kV line to earth
Voltage Dips
IEC61000-4-11
0.5 cycle, 1 cycle,
25 (50Hz)/30 (60Hz) cycles
Voltage Interruptions
IEC61000-4-11
250 cycle (50Hz),
300 cycle (60Hz)
Power frequency magnetic
fi eld IEC61000-4-8
30 A/m (50/60 Hz)
Conducted RF
IEC61000-4-6
3Vrms: 50 kHz to 80 MHz outside ISM bands
6Vrms: 150kHz to 80 MHz in ISM bands
Radiated RF EM fi elds
IEC61000-4-3
3 V/m @ 80 MHz to 2.7 GHz
Proximity fi elds from RF
wireless communications
equipment IEC61000-4-3
IEC 60601-1-2 Chapter 8.10 Table 9
WARNING
Use of eAdvantage® adjacent to or stacked with other
equipment should be avoided because it could result
in improper operation of the device.
WARNING
Portable RF communication equipment should be used no closer than
30 cm (12 inches) to any part of the eAdvantage®.
WARNING
Use of accessories, transducers and cables other than those specifi ed or
provided by the manufacturer of eAdvantage® could result in increased
electromagnetic emissions or decreased electromagnetic immunity of
eAdvantage® and result in improper operation.