33
eAdvantage User Manual (15PL1033 Rev. 7, July 2022)
Symbol Name
Cell
#
Priority
Alarm
delay
On-Screen
Visual Alarm
Audible Alarm
Low O
2
% @
6% volume
lower than
setting
D
Medium
or highbelow
27% of O
2
%
After 3
consecutive
breaths
Follow the
visual alarm logic
depending on the
alarm priority
(Medium) 1 Burst with
3 pulses each, repeat
every 20 seconds
(high) 1 Burst with
10 pulses each, repeat
every 7.5 seconds
High O
2
% @
6% volume
higher than
setting
D Medium
After 3
consecutive
breaths
Solid symbol fl ashing
down with yellow
warning symbol
at 0.7 Hz with
50% usage
1 Burst with 3 pulses
each, repeat every
20 seconds
LEAK O
2
VALVE
O
2
valve
leak during
standby
D High
5 seconds
after turning
the device
ON or the last
demanded
breath
Solid symbol
with fl ashing red
warning symbol
at 1.4 Hz with
50% usage
1 Burst with
10 pulses each, repeat
every 7.5 seconds
LEAK N
2
O
VALVE
N
2
O valve
leak during
standby
LEAK O
2
&
N
2
O VALVES
O
2
& N
2
O
valves leak
during
standby
7.2 Flow Valve Leak Alarm
7.3 Battery Indicators
The eAdvantage® is equipped with fl ow valve leak detection when gas
is supplied to the unit in either the ON and OFF positions.
When the device is ON, it will monitor the output from the 2 fl ow
sensors during standby mode or 5 seconds after the last demanded
breath. Should either of the fl ow sensors detect any fl ow the device
will display the leak alarms on the screen.
Battery status will be displayed on the screen. There are 2 di erent
status indicators, one showing discharge status and the other showing
the charging status. Capacities and alarms for charging and discharging
are as follows:
Note: This feature will be disabled when both input gasses are OFF
(0 PSI).