41
eAdvantage User Manual (15PL1033 Rev. 7, July 2022)
Note: Upon disconnecting AC Power supply, Mixer will automatically
switch to battery operation without a ecting mixer behavior.
AC/DC Power Supply
MODEL PMP90-13-2 01CV0104
INPUT VOLTAGE 100 - 240 VAC
INPUT CURRENT 1.06 - 0.45 A
INPUT FREQUENCY 47 - 63 Hz
OUTPUT VOLTAGE 19 VDC
OUTPUT CURRENT 4.74 A Maximum
DC OUTPUT PLUG 2.5 x 5.5 x 11 mm
DC OUTPUT CABLE LENGTH 6’
AC POWER CORD LENGTH 6’
WEIGHT 642 g
9.4 Electromagnetic Compatibility
The eAdvantage® has been tested and complies with IEC 60601-1-2:2014
requirements and is intended for use in hospital environment except near
active HF surgical equipment where the intensity of electromagnetic
disturbance is high. The user of eAdvantage® should ensure that it is not
used in environments outside those specifi ed below:
Electromagnetic Emissions
Emission test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The eAdvantage® uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The eAdvantage® external power supply is
suitable for use in all establishments other than
domestic and those directlyly connected to the
public Low-voltage power supply network that
supplies buildings used for domestic purposes.
If it is used in a residential environment,
the eAdvantage® might not o er adequate
protection to RF communication services. The
user might need to take mitigation measures,
such as relocation or re-orienting the device.
Harmonic
emissions
IEC61000-3-2
Class A
Voltage fl uctuations/
fl icker emissions
IEC61000-3-3
Complies