4
eAdvantage User Manual (15PL1033 Rev. 7, July 2022)
1. SAFETY
WARNINGS
• Federal Law restricts this device to sale by or on the order of a
physician.
• The eAdvantage® shall only be used by qualifi ed healthcare
professionals trained in its use for the purposes specifi ed under
“Intended Use”.
• The eAdvantage® system is designed for patient self-administration.
• Never attach the face mask to the patient using a head harness.
• The eAdvantage® shall not be used in oxygen rich nor fl ammable
gases or anaesthetic agents environments. Keep away from open
fl ames, sparks and grease/oil.
• It is recommended to use cylinders that are at least 1/4 full. Always
turn ON the cylinder valve slowly and fully.
• The nitrous oxide cylinder should be operated in an upright position.
If the nitrous oxide cylinder is in a valve-down position while the post
valve is open, liquid may be expelled through the vent passages. This
liquid, Nitrous Oxide, can cause burns by freezing on exposed skin.
• Never allow oil or grease to come into contact with any part of the
cylinders, regulators or eAdvantage® system.
• Always use the checklist to ensure that all components are
reassembled correctly and that all disposable items are replaced.
• Before use on a patient, the settings of the delivered gas should be
checked to ensure they are in accordance with the intended use.
• It is recommended to use an oxygen monitor complying with ISO
80601-2-55 whenever the gas mixer is in use by connecting it to the
gas outlet on the device between the device and the circuit.
• This device must be used with the O-Two™ Breathing Circuit,
01CV8037. It is recommended to connect the expiratory tubing of
the circuit to an Anaesthetic Gas Scavenging System complying with
ISO 80601-2-13
• If the alarm sounds continuously, immediately discontinue use and
shut OFF the gas supply.
• After use, always turn OFF the cylinders to ensure that gas cylinders
with su cient volume are attached before returning the unit to its
normal storage position.
• After use always turn OFF the device.
• Do not disassemble any part of the unit except where described
in this manual as any unauthorized disassembly will invalidate the
warranty.