eAdvantage User Manual (15PL1033 Rev. 7, July 2022)
Level II service
The device shall be returned to the manufacturer or a service center
authorized by the manufacturer for level II service every 2 years.
Manufacturer full service
The device shall be returned to O-Two™ Medical Technologies for
manufacturer full Service every 6 years.
Description Procedure Criterion Schedule Factor:
PM
Visual
inspection
User Manual
Chapter 8.3
Monthly
checking
Device in working
order, gas tanks are full,
no missing item
Monthly User
PM Leak test
User Manual
Section 4.3,
# 4.3.1 ~ 4.3.2
No leak
observed
Every 6
months
User
PM Function Check
User Manual
Section 4.3,
# 4.3.3 ~ 4.3.7
Within
specifi cation
Every 6
months
User
Servicing
Level II
service
Service manual
Meet product
specifi cations
Every
2 years
Manufacturer/
service center
Servicing Full service Service manual
Meet product
specifi cations
Every
6 years
Manufacturer
Note: The manufacturer will make available, on request, factory trained
service personnel, circuit diagrams, component part lists, description,
calibration instructions, or other information required to undertake
service and repair of the device.
Note: Single patient use circuit shall be disposed of in accordance with
local protocols.
At the end of their expected service life, the device, power supply, cables
and battery shall be disposed of in accordance with local protocols.