EasyManua.ls Logo

Oakton TEMP 9000 - Auto-Calibration Gateway

Oakton TEMP 9000
25 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10
Alarm Hysteresis
Use the UP & DOWN arrow keys to adjust value.
The user defined value when the alarm will exit an alarm condition. Example: User has a Pro-
cess High alarm programmed with a value of 150°C and has set the Alarm Hysteresis of 1°C.
The alarm has been activated after the temperature reached 160°C. When temperature
reaches 149°C the alarm condition will not be active.
Audible Alarm
Use UP & DOWN arrow keys to change value; ON or OFF.
Turns the audible alarm on or off.
Advanced Menu Gateway
Press the menu key to advance into the advanced setup.
Press the DOWN key to exit to the system status screen.
Auto-Calibration Gateway
Use DOWN arrow to change to Enter Auto-Calibration Setup. Press the MENU key to enter
To skip the Auto-Calibration press the Menu key.
(Skip to Sensor Offset Calibration if calibration isn’t being performed)
Calibration: Step 1 of 6
The use of a secondary calibrated temperature device is required for the calibration.
Follow instruction on the display for all 6 calibration points.