20
Sensor Input: (Model: 89800-01 & 89800-02)
1. Thermocouple (grounded or non-grounded)
2. Automatic cold junction compensation and break protection for sensor.
3. Temperature display: °F, °C, K, °Re, and °Ra display; user selectable (preset for degrees
C)
Range Type J:
-190°C to 1000°C (-310°F to 1832°F)
Range Type K:
-200°C to 1372°C (-328°F to 2502°F)
Range Type N:
-200°C to 1300°C (-328°F to 2372°F)
Range Type R:
0°C to 1768°C (32°F to 3214°F)
Range Type S:
0°C to 1768°C (32°F to 3214°F)
Range Type T:
-200°C to 400°C (-200°F to 752°F)
Range Type B:
200°C to 1800°C (392°F to 3272°F)
Range Type E:
-200°C to 1000°C (-328°F to 1832°F)
Sensor Accuracy:
Thermocouples
Calibration accuracy +/-0.1% of span, +/-
1
o
C, types J, K, T, E, N. Types B, R,
S: ±0.2% of span
Accuracy Span :1000
o
F (540
o
C) min.