12 Practical Guide for operating the C-Quant
46 / 56 Instruction Manual C-Quant ( G/80000/1007/en)
Example 5: "Range" setting too low
There are too many 0 and not enough 1 responses for a reliable estimation of the
straylight value.
The measurement reliability can be improved by repeating the measurement at a
higher "Range" setting (which will be “G” in this case).