15 Cleaning and Care
50 / 56 Instruction Manual C-Quant ( G/80000/1007/en)
15 Cleaning and Care
Cleaning and Disinfection
Attention
Always disconnect the power supply before cleaning the unit.
Î Cleaning and Disinfection
Î Always take care to observe the product descriptions and instruction manuals of
agents and equipment which you use for the care, cleaning and disinfection of the
unit or its accessories.
Î Do not use agents which are aggressive, contain chlorine, are abrasive, or may attack
the finish of the unit.
Dust protection
Î To prevent dust from entering the viewing hole always keep the instrument in the
dust cover included in the delivery when the instrument is not in use.
Cleaning enameled surfaces
Î Take care not to let cleansers enter the instrument.
We generally recommend using an antistatic cleanser for plastic surfaces (to inhibit
recontamination).
Otherwise you can simply use a damp cloth to clean the outer enameled surfaces.
Use a mixture containing equal parts of spirit and distilled water to remove any res-
idues. You can also add a few drops of a commercial detergent to this mixture.
The lens of the viewing hole can be cleaned with a soft cloth or lens brush using al-
cohol or an optical cleanser.
Disinfection
Î After every use disinfect eyepiece, hand rests, keys and other surfaces which have
been in contact with the patient.
OCULUS recommends for disinfection of all surfaces except those of Plexiglass:
Merz+Co, PURSEPTâ-A Disinfection cloth
D-60318 Frankfurt, Germany
Tel.: +49 69 1503 1
Fax: +49 69 596 21 50
Internet: www.merz.de