What to do if OHAUS Scales unit will not turn on?
- TtiffanyleeAug 1, 2025
If the OHAUS scale does not turn on, the power adapter may not be plugged in or properly connected to the scale. Check the power adapter connections to ensure they are secure.
What to do if OHAUS Scales unit will not turn on?
If the OHAUS scale does not turn on, the power adapter may not be plugged in or properly connected to the scale. Check the power adapter connections to ensure they are secure.
What to do if OHAUS TP200S readings are unstable?
Unstable readings on the OHAUS scale can be caused by vibration on the table surface. Place the scale on a stable surface or change the averaging level.
Why OHAUS Scales showing incorrect weight reading?
An incorrect weight reading may occur if the OHAUS scale was not re-zeroed before weighing or if the scale is not properly calibrated. Press the TARE button with no weight on the pan, then weigh the item again. If the issue persists, recalibrate the scale correctly.
What to do if OHAUS Scales displays random segments or locks up?
If the OHAUS scale displays random segments or the display locks up, it could be due to the microprocessor locking up. Turn the scale off, and then turn it on again. If the condition persists, the unit requires servicing.
Why OHAUS TP200S Scales RS232 interface not working?
If the RS232 interface is not working, the print menu settings might not be properly set up, or there may be an issue with the cable connections. Verify that the interface settings in the Print menu correspond to those of the peripheral device. Also, check the cable connections to ensure they are secure.
How to store menu settings or changes on OHAUS TP200S?
To store menu settings or changes on the OHAUS scale, you MUST use the End function to exit the menus and save the settings.
Why OHAUS TP200S Scales cannot display weight in desired unit or access desired weighing mode?
If the OHAUS scale cannot display weight in the desired unit or access the desired weighing mode, it is possible that the desired unit/mode is not set to ON in the Unit Selection of the Setup menu. See Unit Selection section of Setup menu.
Capacity | 200 g |
---|---|
Power Supply | 4 AA (LR6) Batteries (Not Included) |
Display | LCD |
Interface | RS232 |
Pan Size | 120 mm |
Calibration | External |
Units | g, oz |
Describes the different weighing units available for selection.
Details the balance's ability to communicate via RS-232.
Explains the dual range feature on specific models.
Lists all items included in the balance packaging.
Covers environment, leveling, and component installation.
Instructions for installing draft shield and pans.
Guide on connecting the AC power adapter.
Step-by-step guide to powering on the balance.
How to perform weighing and tare operations.
Detailed procedure for using the parts counting feature.
Guide for performing percent weighing measurements.
Used to calibrate the balance for span or linearity.
Adapts balance to environmental conditions and selects settings.
Enables/disables features and programs parameters.
Configures RS-232 interface and printing functions.
Prevents unauthorized access to balance menus.
Details RS-232 hardware, output formats, and commands.
Guides for diagnosing and resolving balance issues.
Technical specifications and limited warranty information.