EasyManuals Logo

Oldham MX 32 Installation And Operating Manual

Oldham MX 32
48 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #25 background imageLoading...
Page #25 background image
4 Operating instructions
21
Defining the type of sensor
To proceed directly to the next step,
press (MENU/ENTER).
Press (-) or (+) to define the type of
sensor connected.
- BRIDGE: sensor (Wheatstone bridge
output).
- EHP: explosive sensor 4-20 mA.
- FIRE: fire detector (ionic, optic, etc.).
- O2: oxygen detector 4-20 mA.
- OTH: another type of sensor (flame,
etc.).
- MAN: see next step.
Confirm by pressing (MENU/ENTER).
Defining the calibration mode for the
sensor
To proceed directly to the next step,
press (MENU/ENTER).
The MAN indicator blinks.
Press (-) or (+) to define the
operating mode for the yellow sensor LED:
- MAN: no visualization (blinking yellow
LED) by the MX 32 controller, of the
sensor being calibrated (sensor
maintenance mode).
- OTH: visualization (blinking yellow
LED) on the controller when the sensor
is being calibrated.
Confirm by pressing (MENU/ENTER).
Defining the time delay after calibration
mode
(displaying a time)
Press (-) or (+) to define the time
delay (in minutes). Confirm the 4 digits by
pressing (MENU/ENTER)
Saving programming
Press:
- (+) to select Yes. The programming
completed will be used.
- (-) to select No. The changes will be
abandoned and the old programming
will be saved.
Confirm by pressing (MENU/ENTER).

Other manuals for Oldham MX 32

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Oldham MX 32 and is the answer not in the manual?

Oldham MX 32 Specifications

General IconGeneral
BrandOldham
ModelMX 32
CategoryController
LanguageEnglish

Related product manuals