Teardown, Cleaning, and Reassembly of the Miscellaneous Parts of the Olympus BH-2 Microscope Frames Revision 1 Page 19 of 45
Figure 56 Thoroughly clean slide-mount base (old style)
The older style nosepiece slide mount is now ready to
be reinstalled onto the stand and put back into service
(see Figure 57).
Figure 57 Slide-mount parts ready for use (old style)
Servicing the New-Style Slide Mount (BHS/BHT)
This section applies only to the newer style assembly
(see Figure 49). The older style assembly is covered in
the previous section.
Disassemble the New-Style Slide Mount (BHS/BHT)
Using a suitable slotted screwdriver, loosen and remove
the stop screw for the nosepiece slide-mount locking
block (see Figure 58).
Figure 58 Remove the locking-block stop screw
Using a suitable slotted screwdriver, loosen and remove
the pivot screw for the nosepiece slide-mount locking
block (see Figure 59).
Figure 59 Remove the locking-block pivot screw
Separate the locking block from the slide-mount base
(see Figure 60).
Figure 60 Separate locking block from slide-mount base
Clean Components of New-Style Slide Mount (BHS/BHT)
Using a cotton swab and a suitable solvent (acetone),
thoroughly clean the nosepiece slide-mount base,
paying special attention to the surface upon which the
locking block slides and the dovetail bearing surfaces.
(see Figure 61).
Figure 61 Thoroughly clean slide-mount base (new style)