Teardown, Cleaning, and Reassembly of the Miscellaneous Parts of the Olympus BH-2 Microscope Frames Revision 1 Page 20 of 45
Using a cotton swab and a suitable solvent (acetone),
thoroughly clean the mating surface of the nosepiece
locking block, where it slides on the mount base (see
Figure 62).
Figure 62 Clean the nosepiece locking block
Use a suitable solvent (acetone) to clean any grease or
contaminants from the threads of the nosepiece locking
thumbscrew (see Figure 63).
Figure 63 Clean threads of the locking thumbscrew
Reassemble the New-Style Slide Mount (BHS/BHT)
Position the nosepiece locking block beneath the
nosepiece slide-mount base, making sure the bearing
surface of the locking block is in contact with the mount
base (see Figure 64).
Figure 64 Position locking block under slide-mount base
Align the pivot-screw hole in the nosepiece slide-mount
base with the tapped hole in the nosepiece locking
block and use a suitable slotted screwdriver to reinstall
the pivot screw (see Figure 65).
Figure 65 Reinstall the locking-block pivot screw
Align the stop-screw hole in the nosepiece slide-mount
base with the tapped hole in the nosepiece locking
block and use a suitable slotted screwdriver to reinstall
the stop screw (see Figure 66).
Figure 66 Reinstall the locking-block stop screw
The newer style nosepiece slide mount is now ready to
be reinstalled into the stand and put back into service
(see Figure 67).
Figure 67 Slide-mount parts ready for use (new style)