SpecificationCLV-180
1-7
ISSUE1
Endoscope and light guide One-touch connection
Light-supply cable (MAJ-1411) Connector on the rear-panel
(Light-supply cable for CV connection)
Expansion cable (MAJ-202/972)
System connector on the rear-panel
(compatible with Endo GATE/Endo ALPHA)
4. Connection
Foot switch (MAJ-1391) Foot switch connector on the rear-panel
5. Emergency
indication
Emergency lamp indicator Indicates if the emergency lamp is disconnected or if the
emergency lamp is in use (when the main lamp will not illuminate).
Lit up: Emergency lamp in use (the main lamp will not illuminate).
Blinking: Emergency lamp is disconnected or has been removed.
Setting retention (1) Settings prior to powering off the unit are retained.
Manual / automatic brightness settings
Airflow settings
Light intensity level
High-intensity mode setting
Special observation function display
Lamp life indicator (with the power ON, the indicator may
be reset when the examination lamp is extinguished by
continuously pressing the counter reset switch.)
6. Panel
Switch illumination system
(not including counter reset
switch)
(1) All switches illuminated
(not including counter reset switch)
1. Dimming
On endoscope removal (1) When the endoscope is removed, the light emanating from
the scope connector is dimmed by the dimmer plate
located in the endoscope socket.
2. Disinfection
Exterior
(1) Disinfect with ethanol for disinfection (70 ethyl or
isopropyl alcohol) or with a mild detergent.
3. Alarm
Temperature switch (1) To ensure safety, when the internal temperature rises
above the specified value, the temperature switch turns off
and the electrical current supply to the lamp is terminated.
The unit detects that the temperature switch has shut off,
and sounds an alarm.
4. Electrical
shock
Protection Class 1 unit (3P power supply)
Region
Sold
1. EU/ EFTA
Y
MDD
Class:II a (rule: 11-1)
CE marking: CEO197
2. Japan N
Pharmaceutical Law (Application of the Pharmaceutical law to assist Asian
market response)
7. Safety
5. Applicable
regulations
3. USA Y
FDC laws