TroubleshootingCLV-180
2-12
ISSUE1
Pin 8H
High-brightness SP ScopeHigh brightness SP Scope Dummy
Pin 2H
Pin 5L
Pin 8L
*Happrox. 5 V
NOReplace Scope Sensor
Confirm following status of J8 when Brightness is set to manual and
light quantity from minimum to maximum:
J8
Pin 2 Voltage change minus to plus.
Iris Unit /
Main Board
1
YESReplace Iris Unit
NO Replace Main Board
2-12 Automatic Brightness failure
View field is dark or excessively light
2-12-1 Estimated location of failure
No
Estimated location failure Inspection Method
1
Light Source Cable
CV-180
Confirm connection to system capable of automatic Brightness.
CV-180
Light Source Cable
2
Manual Brightness failure Refer to 2-11
Confirm following status of J8 when scope move to the distance from
vicinity:
J8
Pin 2 Voltage change minus to plus.
3
Main Board
TES Replace Iris Unit
NO:Replace Main Board