EasyManua.ls Logo

Olympus CLV-180 - Pump Malfunction

Olympus CLV-180
81 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TroubleshootingCLV-180
2-8
ISSUE1
2-8 Pump malfunction
No movement of the pump, or no change in air supply volume when switching the pump ON.
2-8-1 Block diagram
2-8-2 Estimated location of failure
No
Estimated location failure Inspection Method
Scope 1
Confirm that scope is compatible GI scope.
*EVIS 100,130, 160, 180 Series
1
Set Pump Switch to "H".
Confirm following status of J15 on Main Board:
J15 on MAIN BOARD
When GI scope is connected
Pin 2: LOW
Pin 5: HIGH
Pin 8: LOW
Pin 9: GND
* HIGH: approx. 5 V
Scope Sensor
2
NO Replace Scope Sensor
1
Verify Harness connection.
Main Board (J12) <--> Front Panel
2
Verify hose connection.
3
Check stuffing of Reverse Stop Valve
Harness/
Tube
4
Verify rubber of Air Supply Mouthpiece.
Pump 1
Confirm that air supply pressure / volume conform standards when
changing air supply setting to L, M, and H.
Air Flow Checker
Max pressure (at H)
53.9 kPa or less (air supply pressure locked)
Volume
L: 0.5 L/min or more (output pressure: set as 19.6-20.6 kPa)
M: 0.68 L/min or more (output pressure: set as 21.6-22.6 kPa)
H: 1 L/min or more (output pressure: set as 26.5-27.5 kPa)

Related product manuals