SpecificationCLV-180
1-3
ISSUE1
only).
(16) Panel settings are retained, and the light source unit will display at the previous settings the next
time the unit is turned on. However filter switch settings are not backed up, and will revert to the
regular filter.
(17) The exterior and the operating panel may be disinfected with ethanol (70 ethyl or isopropyl
alcohol) or with a mild detergent.
(18) This product has passed the OLYMPUS Eco-Product standard (Environmentally-friendly design
standard).
3 Operational Conditions
(1) Applicable video system centers: CV-180, CV-165
(2) Operational environment:
i) Use in a medical facility under the supervision of a medical doctor.
ii) Do not apply this light source directly to the heart.However, it may be used with the heart when
used in combination with a TYPE CF applied part indicated by a mark on the instrument.
iii) The outer casing of the unit must be grounded for safety.
iv) Do not use the light source in a combustible atmosphere.
v) Do not use the light source with the following electronic equipment or endoscopes:
a. Any devices designed to apply electronic treatment of a patient for which safe usage with
the CLV-180 has not yet been confirmed.
b. Any devices not designed to apply electronic treatment of a patient for which safety (e.g.,
leakage currents) has not yet be confirmed.
vi) Power supply:
Voltage: 100 120 V AC
Frequency: 50/60Hz
Input current: 500 VA
Power fluctuation: Within ±10%
Frequency fluctuation: Within ±1 Hz
vii) Environment
a. Operating environment
Ambient temperature: 10 40°C
Relative humidity: 30 85%
Air pressure: 700 1060 hPa
b. Storage environment
Ambient temperature: 25 70°C
Air pressure: 700 1060 hPa