ISSUE1 INTRODUCTION
c) Shield parts and cover screws with toothed washers.
Failure to attach parts in their original configurations, even if it does not impair product
function, poses the risk of noise radiation and reduced electrical safety.
(8) Use specified parts.
The parts and components of this product are designed to operate under certain anticipated
vibration, heat, chemical exposure, and voltage conditions. Always replace parts with those
specified in the parts list.
(9) Always tighten nuts/screws to the specified torque and only use components of the specified
dimensions.
(10) Do not reuse O-rings, E-rings, or packing. Always use new ones.
(11) When reusing components, remove sealing tape or sealing compound, and clean the
components before use.
(12) Check that no screws or nuts are loose.
After Repair
(1) After repair, inform the site manager of the nature of the problems, the cause, the
countermeasures taken, and the parts replaced.
(2) Always notify the site manager if, during repair, liquid disinfectants, cleaners, or alcohol were
used. Inform the site manager that before using the unit, it is necessary to verify the
concentration levels of disinfectants to ensure that they have not been diluted.
(3) Restore the original unit settings as they were recorded before repair. Request the presence of
site manager to verify the settings with you.
Copyright
© 2005 Olympus Medical Systems Corp. All rights reserved.
Unauthorized reproduction or distribution in part or in whole is prohibited.
Trademarks
OLYMPUS is a registered trademark of the Olympus Corporation.
The company names, product names, and proprietary technical terms in this document are the
trade-marks or registered trademarks of their respective owners.