TroubleshootingCLV-180
2-14
ISSUE1
YES Replace MAIN BOARD
NO Replace FRONT PANEL
Front Panel blinks immediately after switching Power ON not accepting
buttons.
4
SCORP SENSOR
THERMO SENSOR
MAIN BOARD
1
Pin of Thermo Sensor not conductive:
Replace Thermo Sensor
Signals from Scope Sensor are not the following when each type of
scope is connected to Scope Socket:
Replace Scope Sensor
Main Board J15
GI Scope GI Scope Dummy
J15
Pin 2L
Pin 5H
Pin 8L
SP ScopeHeat Cover Positioning Jig
Pin 2L
Pin 5L
Pin 8H
High brightness SP ScopeHigh brightness SP Scope Dummy
Pin 2H
Pin 5L
Pin 8L
*Happrox. 5 V
None of the above cases applicable: Replace Main Board