Problem Cause Remedy Page
9. The stage descends by its own
weight.
Or the focus is lost due to
slippage of the coarse focusing
knob.
The tension adjustment ring is
loosened too much.
Tighten the tension adjustment ring to
get proper tension.
14
10. The field of view is not matched
between two eyes.
The interpupillary distance is not
adjusted correctly.
Adjust it correctly.
15
The difference in diopter of two eyes
is not corrected properly.
Correct it properly.
16
Different eyepieces are used for right
side and left side.
Use the same eyepiece for right side
and left side.
29
11. When changing the objective
from lower magnification to
higher magnification, the
objective collides with the
specimen.
The specimen is placed upside
down.
Place the specimen with the cover
glass side facing up.
10
The cover glass is too thick. Use a cover glass with thickness of
0.17 mm.
10
12. The LED illumination does not
light.
The AC adapter or power cord is
disconnected.
Connect the power cord to a power
outlet securely.
32
13. When changing the
magnification, the specimen is
defocused significantly.
The diopter is incorrect. Adjust the diopter of eyepieces
correctly. 16
Repair request
If you cannot solve the problems even though taking actions described in “Troubleshooting”, please contact Olympus
for assistance. Please provide us the following information at that time.
Product name and abbreviation (Ex.: Biological Microscope CX33RTFS2)
Product number
Phenomena