INSTRUCTIONS
INDUSTRIAL VIDEOSCOPE
IPLEX RX series
IPLEX RT series
IV9620RX
IV9635RX
IV9650RX
IV9675RX
IV9420RX
IV9435RX
IV9620RT
IV9635RT
IV9650RT
IV9675RT
IV9420RT
IV9435RT

Table of Contents

Questions and Answers

  • M
    mary71Sep 8, 2025
    What to do if Olympus Analytical Instruments shows 'IMAGER ERROR'?
    • S
      Sharon AguilarSep 8, 2025
      If you encounter an 'IMAGER ERROR' on your Olympus Analytical Instruments, stop the inspection and turn the instrument off, then on again.
  • D
    Darrell SullivanSep 7, 2025
    How to improve image brightness on my Olympus Analytical Instruments?
    • K
      Kelly MackSep 7, 2025
      If the image brightness is not optimum on your Olympus Analytical Instruments, configure the proper setting by performing the [BRT] lever operation. Check if the objective lens on the insertion tube's distal end or the optical adapter is dirty and clean it. Also, ensure that the external monitor is correctly adjusted and the optical adapter is properly attached. If the optical adapter is old, consider replacing it.
  • T
    Tamara GarzaAug 22, 2025
    What to do if I get an 'OVER CURRENT (USB)' error on my Olympus Analytical Instruments?
    • L
      Lisa IbarraAug 22, 2025
      To resolve the 'OVER CURRENT (USB)' error on your Olympus Analytical Instruments, first, power off the instrument and then turn it back on. Ensure you are using a recommended USB flash drive. If the issue persists, disconnect all USB devices except the flash drive, as other connected devices might be causing the overcurrent.
  • M
    Mandy JohnsonSep 4, 2025
    What to do if my Olympus Analytical Instruments says 'NO RECORDING MEDIA'?
    • B
      Brent ChenSep 4, 2025
      If your Olympus Analytical Instruments displays 'NO RECORDING MEDIA', load the USB flash drive and try again. If the problem persists, turn the instrument off, then on again.
  • S
    Sean OwensSep 3, 2025
    What to do if Olympus Analytical Instruments shows 'HIGH TEMPERATURE (DISTAL END)'?
    • H
      Harold AguilarSep 3, 2025
      If you get a 'HIGH TEMPERATURE (DISTAL END)' message on your Olympus Analytical Instruments, immediately pull the insertion tube out of the observation object.
  • S
    Stephanie FitzgeraldSep 2, 2025
    What to do if I get a 'RECORDING MEDIA ERROR' on my Olympus Analytical Instruments?
    • Z
      Zachary LyonsSep 2, 2025
      To resolve a 'RECORDING MEDIA ERROR' on your Olympus Analytical Instruments, format the USB flash drive directly on the instrument.
  • R
    Rachel PerkinsAug 31, 2025
    How to fix a blurry image on my Olympus Analytical Instruments?
    • R
      Rachel PerkinsSep 1, 2025
      If the image from your Olympus Analytical Instruments is not sharp, try wiping the objective lens on the insertion tube's distal end or the optical adapter with a clean piece of gauze or a cotton swab. Ensure the optical adapter is properly attached and that the external monitor is correctly adjusted.
  • K
    keithjohnsonAug 30, 2025
    Why is the operation load for articulation getting heavier on my Olympus Analytical Instruments?
    • R
      rosejenniferAug 30, 2025
      If you are experiencing heavy operation load for articulation with your Olympus Analytical Instruments, the angulation lock might be engaged. Press down the [ANGLE LOCK] lever to release the angulation lock.
  • S
    Samantha MirandaAug 29, 2025
    What to do if Olympus Analytical Instruments shows 'HIGH TEMPERATURE (MAIN UNIT)'?
    • D
      dhigginsAug 29, 2025
      If you see a 'HIGH TEMPERATURE (MAIN UNIT)' message on your Olympus Analytical Instruments, it indicates the main unit is overheating. Stop the inspection, allow the instrument to cool down, and then turn the power back on.
  • D
    Debra BonillaAug 27, 2025
    What to do if Olympus Analytical Instruments shows 'IMAGE CAPTURE ERROR'?
    • B
      Blake NorrisAug 27, 2025
      If you encounter an 'IMAGE CAPTURE ERROR' on your Olympus Analytical Instruments, stop the inspection and turn the instrument off, then turn it on again.

Summary

Important Information-Please Read Before Use

Safety Precautions

General safety guidelines for instrument handling and usage.

General Safety Precautions

Detailed safety guidelines for operating the instrument safely.

Battery Precautions

Safety guidelines and warnings for battery handling and charging.

2 Nomenclature

3 Pre-observation;pre-operation preparation and inspection

3.1 Pre-observation Preparation

Steps to prepare the instrument before use.

Preparing the Power Supply

Instructions for connecting the battery or AC adapter.

3.2 Pre-operation Preparation

Procedures to perform before starting observation.

3.3 Pre-operation;Post-operation Inspections

Checklists for ensuring instrument readiness and status.

4 Basic Operation

4.1 Viewing an Observation Object

Steps to perform observation of the target.

4.4 Recording Images

Instructions for capturing still and movie images.

6 Stereo Measurement Procedures (RX (stereo measurement function) only)

6.1 Measurement Flow

Overview of the steps for stereo measurement.

6.2 Attaching the Stereo Optical Adapter

Instructions for physically connecting the adapter.

7 Storage and Maintenance

8 Troubleshooting

Troubleshooting Guide

Section covering common issues and error messages.

9 Specifications

Olympus IPLEX IV9635RX Specifications

General IconGeneral
ModelIPLEX IV9635RX
CategoryIndustrial Videoscope
Insertion Tube Diameter6.0 mm
Working Length3.5 m
Insertion Tube Length3.5 m
Field of View90°
Light SourceLED
DisplayLCD
Battery Life4 hours
Image SensorCMOS
Bending Section120°

Related product manuals