What to do if Olympus Analytical Instruments shows 'IMAGER ERROR'?
- KKevin HornSep 8, 2025
If you encounter an 'IMAGER ERROR' on your Olympus Analytical Instruments, stop the inspection and turn the instrument off, then on again.

What to do if Olympus Analytical Instruments shows 'IMAGER ERROR'?
If you encounter an 'IMAGER ERROR' on your Olympus Analytical Instruments, stop the inspection and turn the instrument off, then on again.
How to improve image brightness on my Olympus Analytical Instruments?
If the image brightness is not optimum on your Olympus Analytical Instruments, configure the proper setting by performing the [BRT] lever operation. Check if the objective lens on the insertion tube's distal end or the optical adapter is dirty and clean it. Also, ensure that the external monitor is correctly adjusted and the optical adapter is properly attached. If the optical adapter is old, consider replacing it.
What to do if I get an 'OVER CURRENT (USB)' error on my Olympus Analytical Instruments?
To resolve the 'OVER CURRENT (USB)' error on your Olympus Analytical Instruments, first, power off the instrument and then turn it back on. Ensure you are using a recommended USB flash drive. If the issue persists, disconnect all USB devices except the flash drive, as other connected devices might be causing the overcurrent.
What to do if my Olympus Analytical Instruments says 'NO RECORDING MEDIA'?
If your Olympus Analytical Instruments displays 'NO RECORDING MEDIA', load the USB flash drive and try again. If the problem persists, turn the instrument off, then on again.
What to do if Olympus Analytical Instruments shows 'HIGH TEMPERATURE (DISTAL END)'?
If you get a 'HIGH TEMPERATURE (DISTAL END)' message on your Olympus Analytical Instruments, immediately pull the insertion tube out of the observation object.
What to do if I get a 'RECORDING MEDIA ERROR' on my Olympus Analytical Instruments?
To resolve a 'RECORDING MEDIA ERROR' on your Olympus Analytical Instruments, format the USB flash drive directly on the instrument.
How to fix a blurry image on my Olympus Analytical Instruments?
If the image from your Olympus Analytical Instruments is not sharp, try wiping the objective lens on the insertion tube's distal end or the optical adapter with a clean piece of gauze or a cotton swab. Ensure the optical adapter is properly attached and that the external monitor is correctly adjusted.
Why is the operation load for articulation getting heavier on my Olympus Analytical Instruments?
If you are experiencing heavy operation load for articulation with your Olympus Analytical Instruments, the angulation lock might be engaged. Press down the [ANGLE LOCK] lever to release the angulation lock.
What to do if Olympus Analytical Instruments shows 'HIGH TEMPERATURE (MAIN UNIT)'?
If you see a 'HIGH TEMPERATURE (MAIN UNIT)' message on your Olympus Analytical Instruments, it indicates the main unit is overheating. Stop the inspection, allow the instrument to cool down, and then turn the power back on.
What to do if Olympus Analytical Instruments shows 'IMAGE CAPTURE ERROR'?
If you encounter an 'IMAGE CAPTURE ERROR' on your Olympus Analytical Instruments, stop the inspection and turn the instrument off, then turn it on again.
| Brand | Olympus |
|---|---|
| Model | IPLEX IV9675RT |
| Category | Analytical Instruments |
| Language | English |
Explains signal words used throughout the manual for hazards and notes.
Lists general guidelines for safe handling and operation of the instrument.
Details critical safety guidelines for handling and charging batteries.
Covers steps for preparing the power supply before instrument use.
Details essential procedures to perform before operating the instrument.
Lists items to inspect before and after using the instrument.
Guides on inserting the insertion tube and operating angulation for observation.
Covers functions like freezing, zooming, and adjusting image brightness.
Details the procedures for capturing still and movie images.
Guides on viewing recorded images in full-screen or thumbnail views.
Outlines the step-by-step sequence for performing stereo measurements.
Details the correct procedure for mounting the stereo optical adapter.
Explains how to add or update stereo optical adapter information in the main unit.
Verifies the accuracy of measurement values before and after operations.
Covers selecting measurement methods and specifying measurement points.
Provides solutions for common error messages and operational problems.