Preparation for Use
MOBILE WORKSTATION WM-DP1
10
Some system installations, for instance those using two mobile workstations, may require two
separation transformers to be fitted.
NOTE
The total impedance of the protective earth path for a system may be up
to 0.4
Ω
or higher if the conditions of clause 18 g) of EN IEC 60601-1 are
satisfied.
The power outlets should only be used for powering equipment which
forms part of the endoscopy imaging system.
All equipment forming part of the system should be powered from the
transformer, unless appropriate isolation is employed as described in
Annex BBB of EN IEC 60601-1-1:2001. This particularly applies where
ancillary equipment as described in (i) forming part of the system is
remotely situated.
CAUTION
Failure to observe either of the above mentioned points could result in
the isolation of the entire system being compromised, defeating the
protection provided by the transformer in a single fault condition.
Additional fixings are provided (Figure 4.1) to facilitate shelf adjustment or the fitting of the optional
drawer pack(s).