Troubleshooting
MOBILE WORKSTATION WM-DP1
33
Chapter 11 Troubleshooting
If the problem cannot be resolved, contact Olympus for service. There are no user-serviceable
parts inside the separation transformer. This equipment must not be modified without written
authorisation from Olympus.
Problem Possible cause Remedy
No power to separation
transformer
Separation transformer not
connected to AC wall outlet
Connect transformer cable to
AC wall outlet
No power at AC wall outlet Check power supply to AC wall
outlet
Output fuse blows Item of equipment connected
to transformer is faulty
Disconnect transformer from
power supply. Locate and
repair faulty equipment.
Replace output fuse and
reconnect all equipment.
Circuit breaker(s) ‘trip’ Combined rating of equipment
connected to power sockets
exceeds the maximum VA
rating of the transformer.
Check combined rating is
within the maximum limit,
disconnect equipment if
necessary. Reset circuit
breaker(s) by pressing in
circuit breaker button.
Interference to equipment
connected to the transformer.
Magnetic - proximity of
sensitive equipment to
transformer magnetic field.
Relocate sensitive equipment
to a higher shelf.
Electrical - typically distortion /
interference on monitor display
when equipment is operated.
Check system for damaged
cables or connections.
Electromagnetic - interference
from other systems.
As above, plus investigate
source of interference.