3-13
Using the Infrared Thermometer/Transmitter
3
• To set the low alarm value (OS552A thru OS555A):
1. Press and hold the  key until the Low Alarm Display
Mode (LAL) appears.
2. Press the  key to increment the low alarm value. Press
the  key to decrement the low alarm value.
3. Press the  key to enable the low alarm function. The
icon appears.
If the temperature drops below the low alarm
setpoint, you will hear a beep and the  icon on
the display flashes. The low alarm output voltage
goes high. The alarm LED turns on.
4. To disable the low alarm, press the  key again, and
the  icon disappears.
If you are not in Low Alarm Display Mode (LAL) when
the low alarm goes off, you must press 
the
key to get into the 
Low
Alarm Display Mode.
Then press the  key to disable the low alarm.
The unit maintains its low alarm set point value even
when the power is removed.
Default values of Low alarm set points are: 
OS551A, 552A, 553A  -10°F
OS554A 0°F 
OS555A 1000°F
NOTE
NOTE