GUIDANCE AND MANUFACTURER’S DECLARATION
(Table 2)
Guidance and manufacturer’s declaration electromagnetic immunity
Immunity test
IEC 60601
Test level
Compliance
level
Electromagnetic environment
guidance
OMRON BPM is intended for use in the electromagnetic environment
specified below. The customer or the user of this OMRON BPM should assure
that it is used in such environment.
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Floor should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %.
Not Applicable.
Not Applicable.
±6 kV contact
±8 kV air
±2 kV for
power
supply lines
±1 kV for
input/output
lines
Not
Applicable.
±1 kV
line(s) to
line(s)
±2 kV
line(s) to
earth
Not
Applicable.
±6 kV contact
±8 kV air