SYMBOL CAUSE CORRECTION
Air.plug.disconnected
Insert.the.plug.securely.Refer.to.“Applying.the.
Arm.Cuff”
Arm.cuff.not.applied.correctly
Apply.the.arm.cuff.correctly.Refer.to.“Applying.
the.Arm.Cuff”
Air.is.leaking.from.the.arm.cuff Replace.the.cuff.with.the.new.one
Cuff.over.or.under.inflated
Repeat.measurement.Remain.still.and.do.not.
talk.during.measurement.Refer.to.“Taking.a.
Measurement”
If.“E2”.appears.repeatedly,.inflate.the.cuff.manually.
until.it.is.30.to.40.mmHg.above.your.previous.
measurement.result.Refer.to.“Taking.a.Measurement”
The.arm.cuff.was.inflated.above.299.
mmHg.when.inflating.the.cuff.manually
Do.not.inflate.the.cuff.above.299.mmHg.Refer.to.
“Taking.a.Measurement”
Movement.during.measurement
Repeat.measurement.Remain.still.and.do.not.
talk.during.measurement.Refer.to.“Taking.a.
Measurement”
Clothing.is.interfering.with.the.
arm.cuff
Remove.any.clothing.interfering.with.the.arm.cuff.
Refer.to.“Applying.the.Arm.Cuff”
Device.error Contact.Customer.Service
Sensor.may.be.out.of.alignment Contact.Customer.Service.for.repair.information
Cuff.is.not.applied.correctly Refer.to.“Applying..the.Arm.Cuff”
ERROR INDICATORS AND TROUBLESHOOTING TIPS