appears or a
measurement cannot
be complete after the
arm cuff inflates.
You move or talk during a
measurement and the arm
cuff does not inflate
sufficiently.
Remain still and do not talk
during a measurement. If
“E2” appears repeatedly,
inflate the arm cuff manually
until the systolic pressure is
30 to 40 mmHg above your
previous readings. Refer to
sub-section 3.1.
The systolic pressure is
above 210 mmHg and a
measurement cannot
be taken.
Inflate the arm cuff manually
until the systolic pressure is
30 to 40 mmHg above your
previous readings. Refer to
sub-section 3.1.
appears
The arm cuff is inflated
exceeding the maximum
allowable pressure.
Do not touch the arm
cuff and/or bend the air
tube while taking a
measurement.
If inflating the arm cuff
manually, refer to the end
of sub-section 3.1.
appears
You move or talk during a
measurement. Vibrations
disrupt a measurement.
Remain still and do not talk
during a measurement.
Display/Problem Possible Cause Solution

Questions and Answers: