EN45
11. Error Messages and Troubleshooting
If any of the below problems occur during a blood pressure measurement and an ECG recording, check to make sure that the
monitor is 30 cm from other electrical devices, except for the smartphone that is being used with the monitor. For example,
electrical outlets, PCs, PC monitors or other wireless devices such as Wi-Fi routers which emit radio frequencies. If the
problem persists, refer to the table below.
Monitor and blood pressure measurements
BP reading display /
Problem
Possible Cause Solution
appears
or
the arm cuff does not
inflate.
The BP [START/STOP] button was
pressed while the arm cuff is not applied.
Press the BP [START/STOP] button again to turn the monitor off.
After inserting the air plug securely and applying the arm cuff
correctly, press the BP [START/STOP] button.
Air plug is not completely plugged into
the monitor.
Insert the air plug securely.
Refer to sub-section 3.1.
The arm cuff is not applied correctly.
Apply the arm cuff correctly, then take another measurement.
Refer to sub-section 3.1.
Air is leaking from the arm cuff.
Replace the arm cuff with a new one.
Contact your OMRON retail outlet or distributor.