EasyManua.ls Logo

OneTouch Select Plus Flex - System Accuracy; System Accuracy Results for Glucoseconcentrations <5.55 Mmol;L

OneTouch Select Plus Flex
125 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
105
Detailed information about your system
9
System Accuracy
Diabetes experts have suggested that glucose meters
should agree within 0.83mmol/L of a laboratory
method when the glucose concentration is lower than
5.55mmol/L, and within 15% of a laboratory method
when the glucose concentration is 5.55mmol/L or higher.
Samples from 100 patients were tested using both the
OneTouchSelect Plus Flex™ System and the YSI 2300
Glucose Analyzer laboratory instrument.
System Accuracy Results for Glucose
Concentrations <5.55mmol/L
Percent (and number) of meter results that match the
laboratory test
Within
±0.28mmol/L
Within
±0.56mmol/L
Within
±0.83mmol/L
67.2%
(121/180)
93.3%
(168/180)
96.7%
(174/180)
Y
ellow= 5mm Red=7mm mar
g
in
PF3130451Rev1_OTSPF_OB_I_GB_en_zug_R2.indd 105 3/12/15 11:36 AM
Statement of Use: Verify status before each use
Effectivity Date: Mar 18 2015 Status: Current State: Released
Type: Project File Name: 3130451 Revision: 1

Table of Contents

Other manuals for OneTouch Select Plus Flex

Related product manuals