Cleaning and disinfection
Cleaning and disinfection are different and both should 
be performed. Cleaning is part of your normal care 
and maintenance and should be performed prior to 
disinfection, but cleaning does not kill germs. Disinfection 
is the only way to reduce your exposure to disease. For 
cleaning information, see page80 and for disinfecting 
information, see page82.
Storing your system
Store your meter, test strips, control solution and other 
items in your carrying case. Keep in a cool, dry place 
between 5°C and 30°C. Do Not refrigerate. Keep all items 
away from direct sunlight and heat. 
Cleaning your meter, lancing device 
and cap
The meter, lancing device and cap should be cleaned 
whenever they are visibly dirty and before disinfection. 
Clean your meter at least once per week. For cleaning 
obtain regular strength liquid dish soap and a soft cloth. 
Prepare a mild detergent solution by stirring 2.5mL of 
regular strength liquid dish soap into 250mL of water. 
in